Operator‘s Manual FERITSCOPE
®
FMP30
Page 113
Measuring
The matrix measuring mode is suited for
-ferrite content measurements
when the measurements are to be made on different specimens of the same
type in succession, always at certain reference areas, and the measure-
ments of corresponding areas are to be combined into blocks. The matrix
measuring mode is used, for example, in the chemical industry or in the
boiler and steel construction industries.
Automatic block creation is not available in the matrix measuring mode.
Correspondingly, it is not possible to set a block size for automatic block
creation during the application-specific settings using
MENU
(
5.6 ‘Application Specific Settings’, beginning on Page 44).
7.8.4
Making Measurements in the Matrix Measuring Mode
In the matrix measuring mode, it is possible to change the blocks, i.e., to
select the block where the next reading is to be stored manually or automat-
ically (
10.6.3 ‘Matrix Mode’, beginning on Page 169).
All three Applications contain 20 blocks each. With 3 Applications each
with 20 blocks, a maximum of n=318 readings can be stored per block.
When selecting a reference area where measurements in an Application
are to be made, it should be observed that the same normalization and
corrective calibration is underlying the determination of the
-
ferrite con-
tent. Thus, only such reference areas should be selected for which no other
normalization or corrective calibration is required.
The number of applications and blocks cannot be changed later on with-
out a re-initialization.
On the display, the Matrix measuring mode is indicated by
.
appears on the display when the Matrix measuring mode is enabled.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...