Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
6.3.3
Deleting a Corrective Calibration
When deleting the corrective calibration, only the corrective calibra-
tion of the open Application will be deleted. The corrective calibra-
tions of the other Applications are retained. However, if the linking
mode is enabled (
on the display), the corrective calibrations
of all Applications that are linked with the open Application will be
deleted.
Key sequ. /
Action
Detail of the display Explanation
CAL
Use
CAL
to start the corrective calibration
of the open Application.
[Delete cal: DEL]
: Using
DEL
deletes the
corrective calibration.
[Cancel: CAL]
: Using
CAL
cancels the cor-
rective calibration (the stored corrective cal-
ibration is retained unchanged).
DEL
[Yes: DEL]
: The corrective calibration will
be deleted.
[No: ENTER]
: Cancels the delete proce-
dure.
DEL
The corrective calibration will be deleted.
The instrument is again ready to make mea-
surements.