Duplication prohibited
MP101M
Environnement
S.A
3–31
SEPTEMBER 2013
If the span flow rate does not correspond to the flow rate measured by the instrument
(
difference > 1,5 l/min),
repeat the calibration procedure.
Quit the display by pressing down the [
] key, then the [Stop] key.
REMARK :
The flow value read on the reference flowmeter is valid for a defined reference
temperature T
ref
and a reference pressure P
ref
= 1013 hPa.
Thus, it is necessary to convert this flow for the conditions T
atm
and P
atm
in order to
compare it with the flow indicated by the analyzer.
atm
ref
ref
atm
ref
atm
P
T
P
T
Q
Q
)
(
)
(
)
(
)
Manual adjustment of the motor-driven valve :
From
the
« Calibration details »
screen of the
« Flow rate »
sub-menu, it is possible to manually
control the opening and closing of the motor-driven valve in order to calibrate the flow rate faster.
Press down the [Manu] key : the [--], [-], [0], [+] and [++] keys are displayed (screen shown below).
Connect the flow meter to the flexible channel.
NOTE :
The flexible channel is described in Chapter 1, section
Gas transport system, pa
ge 1-6 ,
and shown in (4) of figure 1-4.
If the flow rate read on the flow meter is higher than the reference flow rate, press down the [-] or
[--] (faster) key.
If the flow rate read on the flow meter is lower than the reference flow rate, press down the [+] or
[++](faster) key.
As soon as the flow rate read on the flow meter corresponds to the reference flow rate, press
down the [0] key
the motor driven valve stops.
NOTE :
The flow rate calibration can be carried out in measurement mode : the new calibration
coefficient « F Coef. » is immediately taken into account when the [Memo] key is pressed
down.
Содержание MP101M
Страница 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Страница 18: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 18 Page intentionally left blank...
Страница 23: ...Duplication prohibited MP101M EnvironnementS A 1 5 SEPTEMBER 2013 Figure 1 3 Front face with the door closed...
Страница 40: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 1 22 Figure 1 12 Outline dimensions...
Страница 56: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 2 14 This page is intentionally left blank...
Страница 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Страница 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Страница 136: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 14 Page intentionally left blank...
Страница 140: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 18 Page intentionally left blank...
Страница 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Страница 154: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 32 Page intentionally left blank...
Страница 156: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 2 Page intentionally left blank...
Страница 161: ...Duplication prohibited MP101M EnvironnementS A 5 7 SEPTEMBER 2013 Page intentionally left blank...
Страница 163: ......
Страница 164: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 10 Page intentionally left blank...
Страница 166: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 6 2 Page intentionally left blank...
Страница 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...
Страница 192: ...EnvironnementS A DNP ARM7 board Duplication prohibited NOVEMBER 2009 8 Page intentionally left blank...