Table 3-2 – Multiplexer (MUX) signals
(Acceptable limits on channel 1 to 16 of the multiplexer)
Chan
nel
Abbreviations
(Displays)
Parameters Lower
limit
Normal Upper
limit
01
0 V
Ground
- 50 mV
0 mV
+ 50 mV
02
+ 5 V
Checking of + 5 V power supply
+ 4800 mV + 5000 mV
+ 5200 mV
03
+ 15 V
Checking of + 15 V power supply
+ 1450 mV + 1500 mV
+ 1550 mV
04
*
- 15 V
Checking of - 15 V power supply
+ 1300 mV + 1500 mV
+ 1700 mV
05
**
+ 600 V
Checking of G.M. counter power supply.
+ 500.0 V
000.0 V
600.0 V
+ 700.0 V
06
Ref. 2,5 V
Voltage reference 2,5 V
+ 2485 mV + 2500 mV
+ 2515 mV
07
Ref. HR
Voltage reference of probe humidity probe
+ 3650 mV + 3700 mV
+ 3750 mV
08
Filter T°
Filter temperature measured under the
G.M. counter
+ 100 mV
+ 250 mV
+ 400 mV
09
Int. T°
Internal temperature of the analyzer,
measured on the mother board
+ 100 mV
+ 250 mV
+ 400 mV
10
Up Pr.
Upstream pressure
+ 3880 mV + 4130 mV
+ 4380 mV
11
Down Pr.
Downstrem pressure
+ 3880 mV + 4130 mV
+ 4380 mV
12
Atm. Pr.
Atmospheric pressure
+ 3880 mV + 4130 mV
+ 4380 mV
13
Rel. hyg.
Relative humidity (if RST option « ON »)
+ 780 mV
+ 2350 mV
+ 3910 mV
14
Atm. T°.
Atmospheric temperature (if RST option
« ON »)
-
+ 2750 mV
-
15
Head T°
Head temperature (if RST option « ON »)
-
+ 2750 mV
-
16
Ref A/D
Reference of the A/D converter
+ 2485 mV + 2500 mV
+ 2515 mV
(*)
Channel 04 = The « -15V signal » of the multiplexer is an indicating measurement, but not a direct
measurement of the – 15V voltage really present on the board and which can be measured with a
voltmeter. The tolerance on the real – 15 V is ± 50 mV.
(**) Channel 05 = 600 V effective only when the (GM) counter is supplied.
Содержание MP101M
Страница 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Страница 18: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 18 Page intentionally left blank...
Страница 23: ...Duplication prohibited MP101M EnvironnementS A 1 5 SEPTEMBER 2013 Figure 1 3 Front face with the door closed...
Страница 40: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 1 22 Figure 1 12 Outline dimensions...
Страница 56: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 2 14 This page is intentionally left blank...
Страница 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Страница 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Страница 136: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 14 Page intentionally left blank...
Страница 140: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 18 Page intentionally left blank...
Страница 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Страница 154: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 32 Page intentionally left blank...
Страница 156: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 2 Page intentionally left blank...
Страница 161: ...Duplication prohibited MP101M EnvironnementS A 5 7 SEPTEMBER 2013 Page intentionally left blank...
Страница 163: ......
Страница 164: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 10 Page intentionally left blank...
Страница 166: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 6 2 Page intentionally left blank...
Страница 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...
Страница 192: ...EnvironnementS A DNP ARM7 board Duplication prohibited NOVEMBER 2009 8 Page intentionally left blank...