MAINTENANCE SHEET
Monitor serial No.:
OPERATION SHEET : 4.3.2
Scope : Check of KNF pump assembly
PAGE : 1/2
Periodicity : 1 year
Replacement parts : Maintenance kit of KNF pump, ref. V02-K-0012
Dates
Procedure (see Figure 4–2) :
1/ Stop the measurement cycle and disconnect the pump from the analyzer before any intervention.
2/ Disassembly of a pump head :
–
Mark with a pen line the cover (T), the cylinder head (C) and the pump body (A) to avoid a
incorrect positioning of parts during reassembly.
–
Unscrew the four hexagon socket head screws (B) of fixation and remove the assembly
composed of the cover (T) and the cylinder head (C) of the pump body.
3/ Replacement of the diaphragm :
–
Unscrew the screw (D) and remove the tightening disc (E), the lock washer (L) as well as the
diaphragm (F).
–
Remove the four screws (G) and remove the cover (H) to access the mechanical part of the
pump body.
–
Place the connecting rod (K) in half-stroke position turning the counterweight, then put in place
the new diaphragm (F).
–
Check the cleanliness of the whole parts, and, if necessary, clean them.
–
Replace the disc (E) and the lock washer (L) on the diaphragm and strongly tighten the whole
using the screw (D).
–
Replace the cover (T) and cylinder head (C) assembly by respecting the position of pen line.
Settle regularly and crosswise, using the hexagon socket head screw.
–
Check operation by manually turning the counterweight. There should not be stiff point. Put in
place the cover (H) with the screws (G).
4/ Replacement of check valves and O-rings :
–
Remove the cover (T) from the cylinder head by unscrewing the six screws (S).
–
Remove the screw of the check valve (N), remove and replace the worn lower check valve of
the cylinder head.
–
Check cleanliness of the cylinder head, cover and housings of check valves. In case of
unevenness, scratches or corrosion, parts replacement is necessary.
–
Remove the screw of the check valve (U), remove and replace the worn upper check valve (P).
5/ Putting back in place the pump head : replace the cover (T) with a new gasket (V) and tighten the
six screws (S) regularly and crosswise.
6/ Do the same with the second pump head.
7/ Put back in place the pneumatic connections between the two pump heads.
8/ Re-connect the pump to the analyzer.
Содержание MP101M
Страница 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Страница 18: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 18 Page intentionally left blank...
Страница 23: ...Duplication prohibited MP101M EnvironnementS A 1 5 SEPTEMBER 2013 Figure 1 3 Front face with the door closed...
Страница 40: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 1 22 Figure 1 12 Outline dimensions...
Страница 56: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 2 14 This page is intentionally left blank...
Страница 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Страница 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Страница 136: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 14 Page intentionally left blank...
Страница 140: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 18 Page intentionally left blank...
Страница 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Страница 154: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 4 32 Page intentionally left blank...
Страница 156: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 2 Page intentionally left blank...
Страница 161: ...Duplication prohibited MP101M EnvironnementS A 5 7 SEPTEMBER 2013 Page intentionally left blank...
Страница 163: ......
Страница 164: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 5 10 Page intentionally left blank...
Страница 166: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 6 2 Page intentionally left blank...
Страница 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...
Страница 192: ...EnvironnementS A DNP ARM7 board Duplication prohibited NOVEMBER 2009 8 Page intentionally left blank...