Troubleshooting
Stamolys CA71HY
40
Hauser
9
Troubleshooting
9.1
Troubleshooting instructions
Although the analyzer is not very prone to faults due to its simple design, problems at the measuring
point can, of course, not be completely ruled out.
Possible errors, their causes and their possible remedies are thus listed below.
9.1.1
Error messages
9.1.2
System errors/process errors without messages
Error message
Possible cause
Tests and / or corrective measures
Calibration failed
If a calibration fails, you can enter a new calibration factor manually (CONFIGURATION menu,
"Calibration factor"). Cancel the error message by switching the analyzer off and on briefly.
If the error occurs frequently, you will have to search for the cause.
Air bubbles in system
Start calibration manually (PARAMETER ENTRY, "1. calibration", change the date accordingly,
start measurement) or enter a new calibration factor.
Incorrect concentration of standard Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER
ENTRY", "Calibration solution") or replace the standard.
Reagents contaminated or aged
Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not
change color after max. 10 min, replace the reagents.
Standard dosing defective
Check the valves for contamination, obstructions (visual inspection). Replace valve hoses if
necessary.
Incorrect photometer
Check the setting in the CONFIGURATION menu.
Cell dirty
Not enough light intensity at
receiver, e.g. due to sedimented
particles
– Flush with 12.5% bleaching lye (monthly maintenance)
– When using a CAT 430:
Check the filter.
Incorrect photometer
Incorrect photometer
Check the setting in the CONFIGURATION menu, "Photometer".
No sample
No sample
Establish sample transfer.
Level measurement defective
Check level measurement at the sample collector.
Spillage error
Leak at canisters or hoses
Replace defective components and clean and dry the analyzer or those components affected by the
leak.
No measuring signal
Cell filled with air
Pump sample for 1 minute (SERVICE).
Photometer defective
Inform Service
Electrical connection
Check all electrical connections and make sure the fuses are firmly seated.
Fuse defective
Replace fuse F4 or F5 (semi-delay 0.2 A)
Error
Possible cause
Tests and / or corrective measures
Measured values always
the same
Reagents contaminated
or aged
Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not change color
after max. 10 min, replace the reagents.
No sample, no reagents
Make sure sample and reagents are supplied, check level monitor and clean if necessary
System blocked
Flush with 12.5% bleaching lye (monthly maintenance)
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...