Installation
Stamolys CA71HY
12
Hauser
Adjusting the level monitoring
You have to adjust conductive level monitoring depending on whether you want to connect one,
two or three analyzers to this single collecting vessel.
1.
Depending on the application, mount the right adjusting pin (
9, item 2) or do
not use any adjusting pin.
2.
Pull the marked pipe (item 3) downwards in accordance with your application (1, 2 or 3
analyzers) to set an optimum collecting volume.
3.4
Installation instructions
To install the analyzer at the intended location, proceed as follows:
1.
Mount the analyzer and secure it to a wall using screws (Ø6 mm / 0.24").
For the installation dimensions, please see the previous chapter.
2.
Using a spirit level, check that the cabinet is hanging or standing straight. Only in this way can
any air bubbles present escape from the cell.
3.
Fit the edge covers (with GFR housing only).
4.
Lay the drain pipe for the reaction products. Where possible, use solid pipes (PVC or PE,
internal diameter 1" with 3% incline).
5.
Insert the valve hoses accordingly. These were partly removed from the valves for
transportation. This prevents the hoses from sticking or pressure being applied to a point in the
hose over an extended period.
a0001367
Fig. 8:
One analyzer
1
M 3x12
2
M 3x35 (adjusting pin)
3
Marking 1
a0001366
Fig. 9:
Two analyzers
1
M 3x12
2
M 3x20 (adjusting pin)
3
Marking 2
a0001365
Fig. 10:
Three analyzers
1
M 3x12
3
Marking 3
1
2
3
1
2
3
1
3
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...