Stamolys CA71HY
Wiring
Hauser
17
!
Note!
• Alarm values 1 and 2 do not need to be connected if the process control system sets its own alarm
values at the analog output.
• When using a sample conditioning system: connect terminals 57 and 53 on CA 71 to the
corresponding terminals of the sample conditioning system. For the allocation of these terminals,
please see the Operating Instructions of the sample conditioning system.
• If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement (sample not
ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.
Function
Designation
Terminal
One-channel version
Terminal
Two-channel version
Mains
L
3
3
N
2
2
PE
1)
1)
Brass screw with bolt at top right in the connection compartment (marked with
*
)
1
1
Alarm value 1, channel 1
COM
25
25
NC
24
24
NO
23
23
Alarm value 2, channel 1
COM
28
28
NC
27
27
NO
26
26
Alarm value 1, channel 2
COM
–
13
NC
–
12
NO
–
11
Alarm value 2, channel 2
COM
–
16
NC
–
15
NO
–
14
Fault
COM
19
19
NC
18
18
NO
17
17
Unassigned
COM
22
22
NC
21
21
NO
20
20
Analog output 1
0/4 to 20 mA
+
36
36
–
35
35
Shield
37
37
Analog output 2
0/4 to 20 mA
+
–
39
–
–
38
Shield
–
40
Sample conditioning
remote control
Input
57
57
0 V
53
53
Channel switchover
Input
–
55
0 V
–
53
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...