Stamolys CA71HY
Operation
Hauser
25
Analog output 1
0 to 20 mA /
4 to 20
mA
Selection for the measuring range scope for channel 1. If
the concentration measuring range is 0 to x mg/l, either
4 mA or 0 mA corresponds to 0 mg/l. The end of the
measuring range is the same in both cases at 20 mA.
Analog output 2
Only two-channel version!
Selection for the measuring
range scope for channel 2. The scopes of measuring range
are independent of each other for channel 1 and channel
2 and are determined by the start of measuring range
setting (channel 1 / channel 2) or end of measuring range
setting (channel 1 / channel 2) in the PARAMETER
ENTRY menu.
Alarm value AV 1-1
NO
NC
!
Note!
Changes only take effect after a reset
(power off/on)!
Setting as to whether the contact for alarm value 1,
channel 1 works as an NO or NC contact.
Alarm value AV 2-1
Setting as to whether the contact for alarm value 2,
channel 1 works as an NO or NC contact.
Alarm value AV 1-2
Two-channel version only!
Setting as to whether the contact for alarm value 1,
channel 2 works as an NO or NC contact.
Alarm value AV 2-2
Two-channel version only!
Setting as to whether the contact for alarm value 2,
channel 2 works as an NO or NC contact.
Error contact
Setting as to whether the contact for the fault message
works as an NO or NC contact.
Current date/time
01.01.96 00:00 to 31.12.95 23:59
Setting the system clock.
Format DD.MM.YY hh:mm.
Calibrate offset
Yes / no
Frequency offset
1)
Pressing the
J
key starts blank value measurement for
compensating the reagent's inherent color.
Frequency offset
– 5000 to +5000
0
Manually changing the frequency offset.
1)
1)
Redetermine the frequency offset every time the reagents are changed or the photometer is replaced. To do so, instead of sample connect deionized water to
the sample inlet to determine the frequency offset (blank value). The value is generally between 0 and 10 Hz.
Menu item
Range of adjustment
(factory settings in bold)
Display
Info
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...