Operation
Stamolys CA71HY
22
Hauser
5
Operation
5.1
Display and operating elements
a0001679
Fig. 19:
Display and operating elements
1
LED display (measured value)
2
LC display (measured value + status)
3
Serial interface RS 232
4
Operating keys with indicator LEDs
5.2
Local operation
The operating keys and the integrated indicator LEDs have the following functions:
RS 232
M CE
E
K
1
2
4
3
Key
Key function
Indicator LED function
H
– “Auto measuring” option
– Back to the main menu from all sub-menus
Alarm value 1 exceeded
– Backwards in the sub-menu
horizontal, see Appendix
Alarm value 2 exceeded
V
– Backwards in the main menu (vertical)
– Increase value
Measuring range exceeded
W
– Forwards in the main menu (vertical)
– Reduce value
Measuring range undershot
F
– Select item
– Accept value, move forward in submenu (horizontal)
1))
1)
The number of decimal places can be changed by pressing the
F
key and the
V
or
W
key simultaneously.
Error message must be called up
J
– Selection in the sub-menu
– Select sign +/-
Cleaning interval running
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...