Stamolys CA71HY
Technical data
Hauser
45
10
Technical data
10.1
Input
10.2
Output
10.3
Power supply
10.4
Performance characteristics
Measured variable
N
2
H
4
[
μ
g/l]
Measuring range
1 to 500
μ
g/l
Wave length
430 nm
Reference wave length
565 nm
Output signal
0/4 to 20 mA
Signal on alarm
Contacts: 2 limit contactors (per channel), 1 collection alarm contact
Optionally: end of measurement (with two channel version, display of channel
no. available)
Load
Max. 500
Ω
Data interface
RS 232 C
Load capacity
230 V / 115 V AC max. 2 A, 30 V DC max. 1 A
Supply voltage
115 V AC / 230 V AC ±10%, 50/60 Hz
Power consumption
Approx. 50 VA
Current consumption
Approx. 0.2 A at 230 V
Approx. 0.5 A at 115 V
Fuses
1 x slow-blow 0.5 A for analyzer electronics
2 x semi-delay 0.2 A for photometer
Time between two measurements
t
meas
= reaction time + rinse time + waiting time + post rinse time + filling
time+delay to reject reagent (min. waiting time = 0 min)
Maximum measured error
±2 % of end of measuring range
Measuring interval
t
meas
... 120 min
Reaction time
• 1 minute
Sample requirement
• 15 ml (0.51 fl.oz.) / measurement
Reagent requirement
1 x 0.4 ml (0.014 fl.oz.)
1.73 l (0.46 gal.) reagent per month with 10 minute measuring interval
Calibration interval
0 to 720 h
Flushing interval
0 to 720 h
Rinse time
Selectable from 20 to 300 s (standard = 60 s)
Post rinse time
30 s
Filling time
20 s
Maintenance interval
6 months (typically)
Servicing requirements
15 minutes / week (typically)
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...