Stamolys CA71HY
Wiring
Hauser
19
4.2.3
Two-channel version
4.3
Switching contacts
One-channel version
Two-channel version
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
Signal outputs
AV 1 - 1
Alarm value 1, channel 1 exceeded or undershot
AV 2 - 1
Alarm value 2, channel 1 exceeded or undershot
AV 1 - 2
Alarm value 1, channel 2 exceeded or undershot
AV 2 - 2
Alarm value 2, channel 2 exceeded or undershot
Fault
Error message, can be retrieved using operation menu
Channel 1/2 or measuring
end
1)
1)
Can be selected alternatively
Displays active channel
Displays "Measurement finished" (5 s)
Analog output
l-1 channel 1
0 or 4 mA = start of measuring range
20 mA = end of measuring range
l-2 channel 2
0 or 4 mA = start of measuring range
20 mA = end of measuring range
Channel selection
Meas. 2
0 V = channel 1
24 V = channel 2
Terminal connection for
condition met
Terminal connection for
condition not met
Terminal connection for
power off
AV 1
A: R:
2525
--
2324
A:
R:
25
25
-
-
24
23
25
-
24
AV 2
A: R:
2828
--
2627
A:
R:
28
28
-
-
27
26
28
-
27
Fault
A: R:
1919
--
1718
A:
R:
19
19
-
-
18
17
19
-
18
Unassigned
2216
13
---
2014
11
22
16
13
-
-
-
21
15
12
22
16
13
-
-
-
21
15
12
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1 - 1
A:
R:
25
25
-
-
23
24
A:
R:
25
25
-
-
24
23
25
-
24
AV 1 - 2
A:
R:
13
13
-
-
11
12
A:
R:
13
13
-
-
12
11
13
-
12
AV 2 - 1
A:
R:
28
28
-
-
26
27
A:
R:
28
28
-
-
27
26
28
-
27
AV 2 - 2
A:
R:
16
16
-
-
14
15
A:
R:
16
16
-
-
15
14
16
-
15
Fault
A:
R:
19
19
-
-
17
18
A:
R:
19
19
-
-
18
17
19
-
18
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...