Stamolys CA71HY
Troubleshooting
Hauser
41
Measured values
inaccurate
Incorrect concentration
of standard
Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER ENTRY",
"Calibration solution") or replace the standard.
Reagents contaminated
or aged
Simple check: Mix 5-10 ml of standard solution and 5 ml of reagent in a beaker. If it does not change color
after max. 10 min, replace the reagents.
Reagent blank value too
high
After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,
"Offset calibration")
Wrong dimension
Check the setting in the CONFIGURATION menu, "Measuring unit" (e.g. ion instead of element).
Wrong cell
Check the setting in the CONFIGURATION menu, "Photometer"
Sample suction time too
short
Increase the suction time (CONFIGURATION, "Delay to sample")
Matrix effects
(substances that interfere
with the photometric
method)
Detect interfering substances (see Technical Information, "Measuring principle"),
possibly use sample conditioning
Filter life too long
Take a counter sample at the analyzer inlet and check the concentration in the laboratory. If there is no
deviation to the analyzer measured values, replace the ultrafiltration modules or reversible flow filters more
frequently.
System blocked or
contaminated
Flush with 12.5% bleaching lye (monthly maintenance)
Dosing
Replace pump hoses.
Cell dirty
Flush with 12.5% bleaching lye first and then with 5% hydrochloric acid
Measured values
inaccurate
Standard is dosed to the
sample
Check the valves and valve settings. Replace valve hoses if necessary.
Counter sample in the
laboratory delivers
deviating measured
values
Sample aging
Shorten time between sampling and analysis.
Measured value transfer
defective
Wrong size for analog
output
Check the setting (CONFIGURATION, "Analog output 1" or " 2").
Incorrect measuring
range
Adjust measuring range (PARAMETER ENTRY, "Measuring range")
Background noise
Check electrical cables for interference from strong sources of induction.
Analyzer will not switch
on
No power
Check the electrical connection and ensure power supply.
Fuse
Replace fuse F1 (slow-blow 0.5 A)
Analyzer is running but
display reading is garbled
or off
Initialization failed
Switch the analyzer off and after approx. 30 seconds back on.
Pumps fail to run
Leak
See "Spillage error" error message
Leak sensor bypassed
Interrupt the contact between the two leak sensors (terminals 67-66)
Fuse
Check all fuses and replace if necessary.
Pump defective
Service
Measurement does not
start
Leak at photometer
Service
"Measurement" display
flashes
Time of 1st
measurement not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired
Change parameter setting.
Calibration does not start
Time of 1st calibration
not reached
The date must be between 01.01.1996 and the current date.
Interval not expired or 0
h
Change parameter setting.
Leak at photometer
Service
Error
Possible cause
Tests and / or corrective measures
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...