Troubleshooting
Stamolys CA71HY
42
Hauser
9.2
Spare parts
9.2.1
Overview for
9.2.2
Spare parts for sample and reagent transportation
Flushing does not start
Time of 1st flushing not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired or 0
h
Change parameter setting.
Leak at photometer
Device or discharge
blocked
Remove blockage.
Service
Blockage, deposits in
device
Water hardness
Lime deposits can be removed by flushing with 5% hydrochloric acid.
If necessary, dose EDTA in sample stream to prevent precipitation (EDTA not with CA71HA).
Not enough sample
conditioning
Shorten the cleaning intervals of sample conditioning.
Error
Possible cause
Tests and / or corrective measures
a0006410
Fig. 24:
Overview of components and spare parts
A
B
C
P
Sample input, two-channel version
Standard and reagent canisters
Collecting vessel
Sample
R
R1
S
Cleaner canister
Reagent 1 canister
Standard solution canister
Item
Spare part
Order number
120
Norprene hose, 1.6 mm
51504116
121
C-Flex hose, 3.2 mm (permeate inlet and overflow with sample collector.)
51504114
122
C-Flex hose ID 6.4 mm
51504115
123
C-Flex hose, 1.5 mm
51512535
B
B
C
A
155
131
134
130
141
143
121
135
135
121
170/171
154
R
S
R1
V3
V1
V2
161
160
161
160
V2
V1
120
123
130
144
200
122
142
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...