Stamolys CA71HY
Operation
Hauser
29
5.2.9
DATA MEMORY-calibration data
!
Note!
You get to the data memory calibration data by means of the CONFIGURATION menu. In the
"Calibration factor" menu item, press the
J
key.
5.3
Calibration
!
Note!
The calibration is not performed with a hydrazine standard (short shelf life) but with a color
equivalent.
The first calibration step is an offset measurement of a sample (=basic value). After that, the cell is
rinsed with the color equivalent. Finally, color equivalent and reagent are pumped into the cell and
the value is determined (=plateau value).
5.3.1
Standard calibration data
The signal strength is processed device-internally as a frequency.
The following table provides an overview of the standard calibration data.
!
Note!
Compare these values to your own data.
After changes in the CONFIGURATION menu and in the event of software updates, you can check
and, if necessary, change the calibration data in the sub-menu.
Menu item
Display
Info
Calibration factor
This data memory contains the last 100 calibration factors
with date and time. If no values are available, "Empty set"
appears. Browse through the data sets with the
V
and
W
keys.
Serial output
only possible via PC!
No display
You can output
all
data sets (in ASCII-format) via the
serial interface. For this, the PC must be configured as
follows: 9600, N, 8, 1.
To send data, the PC must send the ASCII character 67
("c").
Clear data
This deletes
all
data sets.
Measuring range
Concentration [
μ
g/l]
Frequency [Hz]
Hydrazine, HY-A
1 to 500
μ
g/l
0
50
100
150
200
250
300
350
400
500
0
37
81
119
153
195
231
269
302
376
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...