BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
17
•
“Error code:
5” = No RTD (Pt100) and amplitude too low (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4)=5)
•
“Error code: 1024” = No pressure sensor connected (bit 10)
•
“Error code: 1029” = No RTD (Pt100), amplitude too low, no pressure
sensor connected (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4], bit 10 [2
N
Value 1024] = 1029)
Graph screen
The oxygen values of the current measurement session are displayed in a
graph; the last measurement value of the
Current Measurement
is shown at
the top of the screen. Refer to Figure 4–20.
Table 4–1
Error codes
Bit
2
N
Value
Error
0
1
No RTD (Pt100)
1
2
No sensor selected
2
4
Amplitude too low
3
8
SD card defect
4
16
Reference amplitude out of range
5
32
Photo diode saturated
6
64
Signal overflow
7
128
Signal overflow
8
256
Reserved
9
512
Critical error. Refer to
10
1024
No pressure sensor / pressure sensor out of range
11
2048
Reserved
12
4096
Storage space full
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...