4-20mA calibration
Use the 4-20mA CALIBRATION screen to calibrate the
Output
and
Input
. The
analyzer is delivered in the calibrated state but can be calibrated to other
devices in your measurement system.
Calibrating the output
Use the following procedure to calibrate the work flow for either Output 1 or
Output 2. Refer to Figure 4–59.
1.
Connect a current measurement device to the respective output.
This will serve as the reference device.
2.
Set the 1st Point value to any low value, e.g., 4.00 mA. The value
will be applied immediately.
a.
Press the Apply button or enter another value.
3.
Read the current value shown on the reference device, e.g.,
3.90 mA.
a.
Use the +/- symbols in the
Adjust
column next to the
1st Point
value to adjust the values accordingly.
Factory calibration will be lost if the analyzer is re-calibrated.
Figure 4–58
Analogues - Concentration alarm
relay
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...