OXY5500 Optical Oxygen Analyzer
Hauser
5
–
13
Compensation temperature
This value is used for compensation of the oxygen calculation.
Measurement interval
The oxygen measurement interval can be set between 1 and 359999. Setting
the interval to “0” will lead to Modbus error response with code 3.
The measurement values may be read any time but are updated only with the
interval set in these registers. Hence, polling the measurement values in a
higher rate than the measurement interval should be omitted as it leads to
unnecessary traffic on the bus.
Device ID RS-485, RS-232 and Ethernet
Sets the device ID used in the Modbus RTU communication. If a value over 32
is set, the device will reset its ID to 1, which may lead to communication errors.
If no ID is set, or the ID is not known, set the ID via broadcast (ID=0).
Table 5–4
Definition of register 2411
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
2411
2
Float: Temperature
value in
°
C
Yes
Table 5–5
Definition of register 3499
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
3499
2
Int32: Interval value in
seconds
Yes
Table 5–6
Definition of register 4095
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
4095
2
Int32: Device ID of RS-
485. Minimum 1, Maxi
-
mum 32
Yes
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...