BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
5
–
5
Table 5-1
Holding registers (Continued)
Register
Name
Address
Size
Variable
Type
Description
Write
Access
Device ID
RS-485
4095
2
Int32
Sets the device ID that is
used in Modbus RTU com
-
munication (range 1-32).
Refer to Table 5–6.
Yes
Device ID
Minimum
RS-485
4097
2
Int32
Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-485
4099
2
Int32
Device ID Address Limit:
Maximum.
No
Baud rate
RS-485
4101
2
Int32
Code for Baud rate where:
3 = 9600
4 = 19200
5 = 38400
6 = 57600
7 = 115200
Yes
Baud rate
Minimum
RS-485
4103
2
Int32
Minimum code for baud
rate.
No
Baud rate
Maximum
RS-485
4105
2
Int32
Maximum code for baud
rate.
No
Parity
RS-485
4107
2
Int32
Parity for the RS-485 out
-
put where:
0x00 = Even Parity
0x01 = Odd Parity
0x02 = No Parity
1
1. When using “No Parity,” an additional Stop-bit will be sent (2 Stop bits).
Yes
Device ID
RS-232
4109
2
Int32
Sets the device ID used in
the Modbus RTU communi
-
cation (range 1 to 32).
Refer to Table 5–7.
Yes
Device ID
Minimum
RS-232
4111
2
Int32
Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-232
4113
2
Int32
Device ID Address Limit:
Maximum.
No
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...