BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
5
–
17
Ethernet IP
Example:
Writing the following bytes:
0x 01 10 16 2D 00 08 10 00 C0 00 00 00 A8 00 00 00 01 00 00 00 0A 00 00 1F B1
will result in the IP 192.168.1.10
In detail:
0x 01
Slave address (int32 “01”)
0x 10
Function Code
0x 16 2D
Starting Address (5677 in int32 representation)
0x 00 08
Quantity of registers
0x 00 C0 00 00
Octet 1 (int32 192)
0x 00 A8 00 00
Octet 2 (int32 168)
0x 00 01 00 00
Octet 3 (int32 1)
0x 00 0A 00 00
Octet 4 (int32 10)
0x 1F B1
CRC16
Hauser recommends deactivating the current
measurement before changing any settings. The device will hold
its last analog output value until the next measurement.
Table 5–15
Definition of register 5677
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5677
8
Int32:
Ethernet IP
Octet 1
Int32:
Ethernet IP
Octet 2
Int32:
Ethernet IP
Octet 3
Int32:
Ethernet IP
Octet 4
Yes
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...