BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
43
3.
Select Yes to stop the measurement in order to move to the
CALIBRATION windows. Use the
buttons to navigate
between input fields.
To enter editing mode
1.
Press
.
a.
Change the setting or value (one digit at a time) by pressing the
buttons.
b.
Make the desired change to an input field.
2.
Press
again to save the changes.
To exit editing mode
1.
Press
to cancel and exit.
Performing pre-calibration
1.
Connect the analyzer to a nitrogen (N
2
) bottle.
2.
Set the flow to 1.5 SLPM.
Figure 4–44
Message window - Stop
measurements during configuration