OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
56
Hauser
The values will be used for calculating the output or input value on the next
measurement.
•
Bilinear: The
High Value
,
Mid Value
and
Low Value
can be entered.
Refer to Figure 4–57. The units are the same as those used in Linear
mode. The values will be used for calculating the output or input
value on the next measurement.
Figure 4–55
Analogues - 4-20mA values for
mode “Off”
Figure 4–56
4-20mA values for mode “Linear”