Appendix A: Specifications
Table A–1
OXY5500 analyzer specifications
Application Data
Target Components
O
2
Principle of Measurement
Fluorescent Quenching
Typical Measurement Ranges
OP-9
1
1. Analyzer ships with one sensor. Additional sensors are optional.
OP-6
1
OP-3
1
0 to 200 ppmv (default)
0-10 to 10-1,000 ppmv
2
User setting
2. Accuracy specification applies to concentrations up to 300 ppmv.
0 to 5%
0-1 to 0-5%
User setting
0-20%
0-10 to 0-20%
User setting
Lower Limit of Detection
0.5 ppmv
20 ppmv
300 ppmv
Accuracy at 20 to 25
°
C
±
5% of Reading
±
3% of Reading
±
2% of Reading
Repeatability
±
1% of Reading
Measurement Update Time
Programmable Sampling Rate (default 30 seconds)
Temperature Range
3
(Configurable)
3. Configurable based on OP-9, OP-6, OP-3 probe constants. Refer to the Calibration Certificate.
1) 0
°
C
to 60
°
C (0
°
F to 140
°
F)
2) -20
°
C
to 50
°
C (-4
°
F to 122
°
F)
Sample Inlet Pressure
140 to 275 KPaG (20 to 40 PSIG) to sample panel regulator
Sample Pressure Range
800 to 1400 mbara
Maximum Probe Pressure
275 KPaG (40 PSIG)
Sample Flow Rate
Typical 1.0 SLPM (2.1 SCFH)
Recommended Calibration
Two-point calibration in oxygen-free environment (nitrogen)
and a second span point (cylinder gas). Validate with O
2
in N
2
reference (cylinder gas).
Electrical & Communications
Input Power
(Voltage and Max. Power)
108 to 253 VAC, 50/60 Hz; 5.3W at 120 VAC; 6.6W at 240 VAC or
9 to 30 VDC (CSA), 18 to 30 VDC (IEC/ATEX); 4.7 W at 24 VDC
Communication
Analog
: Qty (2) 4-20mA sourcing power outputs and (1) 4-20
mA input (sample pressure)
Fieldbus
: RS-232C, RS-485, Ethernet 10/100 with Modbus
Output Relays
: Qty (2) 250 mA max load (Concentration and
Fault Alarms)
USB 2.0 works with Service Software only
4 GB Internal Memory with Internal Data Logging
LCD Display
Concentration, Temperature, Sample Rate, Data Logging, Diag
-
nostics, plus Full Menu for Setup, Calibration, etc.
Service Software
• Windows software.
• Connected via USB port.
• Download data logs, trend and monitor, calibrate and trouble
-
shoot.
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...