Setting the pressure for RATA calculation
After stopping the currently running measurement, the PRESSURE FOR RATA
CALCULATION screen displays. Refer to Figure 4–48.
•
Select Auto and the atmospheric pressure will be measured via the
4-20mA Input.
•
Select Manual if there is no pressure sensor connected to the
analyzer.
a.
Type in the current atmospheric pressure value in the respective
unit (
hPa
,
mbar
,
PSI
,
atm
or
torr
).
b.
Press
to save changes.
Setting the temperature for RATA calculation
•
Select Auto to measure the temperature for RATA calculation with the
RTD probe (Pt100 temperature sensor).
•
Select Manual if the temperature for RATA calculation is known.
Temperature values can be input in
°
C,
°
F or K.
a.
Switch to the desired temperature unit and change the
temperature value in the input field.
b.
Press
to save changes.
Figure 4–48
Pressure for RATA calculation
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...