OXY5500 Optical Oxygen Analyzer
BA02195C
3
–
4
Hauser
Mounting the analyzer
1.
Select a suitable location to mount the analyzer. Choose a shaded
area or use an optional analyzer hood (or equivalent) to minimize
sun exposure.
2.
Locate the mounting holes on your unit. Refer to Figure 3–1 and the
system drawings in Appendix A.
3.
For wall installations, mark the centers of the top mounting holes.
Mounting dimensions are shown in Appendix A.
4.
Drill the appropriate size holes for the screws you are using.
5.
Hold the analyzer in place and fasten with the top screws.
6.
Repeat for the bottom mounting holes.
Once all four screws are tightened, the analyzer should be very secure and
ready for the electrical connections.
Hauser analyzers are designed for operation within the
specified ambient temperature range. Refer to Appendix A. Direct
sun exposure in some areas may cause the analyzer temperature
to exceed the maximum range.
Figure 3–1
Analyzer mounting locations
MOUNTING
HOLES
MOUNTING
HOLES
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...