BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
47
2.
Press
to perform a Relative Accuracy Test Audit (RATA). This
will open a message window with the question, “Measurement
Active. Abort for calibration?” Refer to Figure 4–47.
3.
Select Yes and stop the measurement to move to the CALIBRATION
screen.
4.
Use the
buttons to navigate between input fields.
To enter editing mode
1.
Press
.
2.
Change the setting or value (one digit at a time) by pressing the
buttons.
3.
Make the desired change to an input field.
4.
Press
again to save the changes.
To exit editing mode
1.
Press
to cancel and exit.
Figure 4–47
Message window - Stop
measurements for calibration
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...