BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
33
2.
Press the Next button in the upper right corner of the screen, then
click
.
The display will change to the CALIBRATION DATA screen. Refer to Fig
ure 4–35. If a calibration was performed with a previously connected
sensor, the data from that calibration is shown.
Calibration
Calibration pressure and temperature are set from the CALIBRATION
SETTINGS and CALIBRATION TEMPERATURE screens, as shown below.
Setting calibration pressure
Refer to Figure 4–36 for a view of the CALIBRATION SETTINGS screen. The
following instructions provide setting information.
On the Calibration Certificate (refer to Figure 4–34 on page 4–32),
“T0” is shown under the Calibration Data section, Temperature
column as Cal0 and Cal2nd.
On the Calibration Certificate, “pATM” is shown as “Atmospheric
Pressure” under the Calibration Specifications section
during Cal0
and Cal2nd. Refer to Figure 4–34 on page 4–32.
Figure 4–35
Calibration data screen
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...