OXY5500 Optical Oxygen Analyzer
BA02195C
5
–
10
Hauser
Table 5-1
Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
Cal0
5521
2
Float
Calibration value: Phase
shift of low oxygen calibra
-
tion point (Default: 59.9).
Yes
T0
5523
2
Float
Calibration value: Temper
-
ate at the low oxygen cali
-
bration point in
°
C
(Default: 20.0).
Yes
O2-2nd
5527
2
Float
Calibration value: Oxygen
concentration of the high
oxygen calibration point in
the unit defined in register
5535 (O2-2nd
Unit).
Yes
Cal-2nd
5529
2
Float
Calibration value: Phase
shift of high oxygen cali
-
bration point (Default:
26.3).
Yes
T2nd
5531
2
Float
Calibration value: Tem
-
perature at the high cali
-
bration point in
°
C.
Yes
pATM
5533
2
Float
Calibration value: Pressure
at the high oxygen calibra
-
tion in mbar.
Yes
O2-2nd Unit
5535
2
Int32
The unit for the O2-2nd
value, where:
0x4000.0000 = ppmv
0x0000.0010 = % O
2
Yes
Ethernet
Obtain IP
Mode
5675
2
Int32
Activates or deactivates
DHCP. Entering “1” will
obtain IP automatically.
Yes
Ethernet IP
5677
8
Int32
The Ethernet IP. Each pair
of registers holds one octet
of the address. Refer to
17 for details. This regis
-
ter will only be used if reg
-
ister 5675 is set to “0”
(DHCP off).
Yes
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...