![Endress+Hauser SpectraSensors OXY5500 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/endress-hauser/spectrasensors-oxy5500/spectrasensors-oxy5500_operating-instruction_2407533067.webp)
BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
37
Table 4-4
Calibration materials/equipment (Continued)
Material/
Equipment
Specifications
Vendor; P/N
(if available)
Notes
Tube reducer
Stainless steel tube
fitting, reducer, 1/8 in.
x 1/4 in. tube OD
or
Stainless steel tube
fitting, reducer, 6 mm
x 3 mm tube OD
Swagelok;
SS-200-R-4
SS-6M0-R-3M
(quantity 2)
Port
connector
1/4 TF, OD, 316SS
or
6 mm TF, OD, 316SS
Swagelok;
SS-401-PC
SS-6M1-PC
(quantity 2)
Figure 4–38
General layout for cylinder and
Hauser analyzer connections
CYLINDER VALVE
DUAL STAGE PRESSURE
REGULATOR
SHUT-OFF VALVE
THREE-WAY
BALL VALVE
CAL 0
CAL 2nd
STAINLESS STEEL
TUBING
Port 2
Port 1
C
TO VENT
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...