OXY5500 Optical Oxygen Analyzer
BA02195C
5
–
12
Hauser
Measurement control
This register is used to activate interval measurement and also to trigger a
measurement. It is bit-coded as shown in Table 5–3
.
Table 5-1
Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
LED Intensity 5711
2
Int32
The signal LED intensity.
The allowed range is 0x00
(lowest) to 0x0A (highest)
Yes
Timestamp
8231
2
Int32
This is the current system
time, defined as the num
-
ber of seconds that have
elapsed since 00:00:00,
Thursday, 1 January 1970
(Unix time, ISO8601).
NOTE
: Values below
1493050000 will lead to
error code “illegal value”.
Yes
Table 5–2
Definition of register 5707
Start
Register
Number of
Registers
Reg3 / Reg4
Write
Access
5707
2
Int32:
Bit codes control regis
-
ter. Refer to Table 5–3.
Yes
Table 5–3
Measurement control register bit definition
Bit
Description
0
Interval on/off (delete to turn off, set to turn on)
1
Status-Bit: Set when a measurement is currently performed.
will be deleted after completing the measurement. Setting this
bit does not trigger any action.
2
Start measurement (single scan or continuous)
3 - 31
Reserved
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...