BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
59
4.
Set the
2nd Point
value to any high value, e.g., 20.00 mA. The value
will be applied immediately.
a.
Press the Apply button or enter another value.
5.
Read the current value shown on the reference device, e.g.,
19.54 mA.
a.
Use the +/- symbols in the
Adjust
column next to the
2nd Point
value to adjust the values accordingly.
Example
: The analyzer shows a value of 19.54 mA and the value should be
20.00 mA. Press the
button until the desired value is
registered.
6.
To test the calibration, apply some test points by selecting different
percentage values, such as 0%, 25%, 50%, 75% or 100%, which
corresponds to 4mA, 8mA, 12mA, 16mA and 20mA. Check the
values with the reference device. If satisfied with the calibration,
press the
button to save.
Calibrating the input
The procedure for calibrating the input is similar to the output procedure noted
above. Use the following steps to calibrate the input. Refer to Figure 4–60.
1.
Apply a low current to the OXY5500.
2.
Enter this value in the
Reference
column in the
1st Point
row.
+
Figure 4–59
Analogues - 4-20mA calibration
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...