BA00116D/06/EN/13.10
71123867

Valid as of version
V 3.06.XX (Device software)

Description of Device Functions

Proline t-mass 65
MODBUS RS485

Thermal Mass Flow Measuring System

Содержание Proline t-mass 65 MODBUS RS485

Страница 1: ...BA00116D 06 EN 13 10 71123867 Valid as of version V 3 06 XX Device software Description of Device Functions Proline t mass 65 MODBUS RS485 Thermal Mass Flow Measuring System...

Страница 2: ......

Страница 3: ...NG VALUES 7 4 SYSTEM UNITS 8 5 QUICK SETUP 11 6 OPERATION 23 7 USER INTERFACE 25 8 TOTALIZER 1 2 28 9 HANDLING TOTALIZER 30 10 STATUS INPUT 31 11 COMMUNICATION 33 12 PROCESS PARAMETER 35 13 GAS 39 14...

Страница 4: ...igned to each group You select a group in order to access the individual functions for operating and configuring the measuring device An overview of all the groups available is provided in the table o...

Страница 5: ...R USER INTERFACE 25 ASSIGN LINE 1 ASSIGN LINE 2 100 VALUE LINE 1 100 VALUE LINE 2 FORMAT DISPLAY DAMPING CONTRAST LCD BACKLIGHT TEST DISPLAY TOTALIZER 1 2 28 ASSIGN TOTALIZER SUM OVERFLOW UNIT TOTALIZ...

Страница 6: ...TERNAL HEIGHT INTERNAL WIDTH MOUNTING MOUNTING SET LENGTH INSERTION DEPTH ZERO POINT FLOW CONDITIONER SUPERVISION 51 ACTUAL SYSTEM CONDITION PREVIOUS SYSTEM CONDITION ALARM DELAY SYSTEM RESET OPERATIO...

Страница 7: ...ected volume flow appears on the display The corrected volume flow is calculated from the measured mass flow and the reference density of the gas Display 5 digit floating point number including unit e...

Страница 8: ...FLOW Description For selecting the unit required and displayed for the corrected volume flow The unit you select here is also valid for Low flow cut off The following time units can be selected s sec...

Страница 9: ...al time unit Options US unit 26 tons 27 30 kBtu time unit 31 34 MBtu time unit 35 38 GBtu time unit Factory setting kW or kBtu h country dependent 55 MODBUS register Data type Access 2132 Integer read...

Страница 10: ...ing kg m or lb ft country dependent 55 MODBUS register Data type Access 2107 Integer read write UNIT LENGTH Prerequisite This function is only available for the insertion sensor t mass 65I Description...

Страница 11: ...te This function is only available for insertion sensors t mass 65I Description Use this function to start the application specific setup to calculate the insertion depth for the insertion sensor For...

Страница 12: ...nt parameters can be saved to the T DAT as a backup If the transmitter is replaced for some reason the data from the T DAT can be loaded into the new transmitter EEPROM Options 0 CANCEL 1 SAVE from th...

Страница 13: ...s Heat Flow Unit Heat Flow Unit Quantity Heat Pressure Unit Pressure Totalizer 1 Totalizer 2 Yes No Configure another totalizer Unit Totalizer 1 Assign Totalizer 1 Assign Gas Group Unit Calorific Valu...

Страница 14: ...s required n Only units not yet configured in the current setup are available for selection in each cycle o The YES option remains visible until all the units have been configured NO is the only optio...

Страница 15: ...Setup A0005503 en Fig 2 Quick Setup communication E E Esc E XXX XXX XX Quick Setup HOME POSITION QS Communication Fieldbus Address Baudrate Transmission Mode Parity Byte Order Delay Telegram Reply Set...

Страница 16: ...THICKNESS and OUTER DIAMETER The function INTERNAL DIAMETER displays the calculated internal diameter and is read only RECTANGULAR Enter the INTERNAL HEIGHT INTERNAL WIDTH and WALL THICKNESS of the du...

Страница 17: ...med as one single gas or one gas mixture of up to 8 components A single gas can be selected from a list of standard gases or setup for other suitable types of gases such as Ozone using manual correcti...

Страница 18: ...from the choose list enter the MOLE for each GAS TYPE only if NUMBER OF GASES is 2 and more The error message CHECK VALUES appears if the total mixture does not equal 100 Go back and check the mixture...

Страница 19: ...essures to absolute pressure If a pressure compensating input is being used then the input signal value overrides the manually programmed value The pressure input value applies to both gas groups i e...

Страница 20: ...UP 2 Note If only one gas group is used then leave mode 2 as default settings The units of measure are selected in the SYSTEM UNITS section 8 Auto Gross The gross heating value or higher heating value...

Страница 21: ...bon Monoxide CO 10 1 4 34 10 1 4 34 Hydrogen Sulphide H2S 15 2 6 54 19 49 8 38 Methane CH4 50 02 21 51 55 52 23 87 Ethane C2H6 47 5 20 43 51 93 22 33 Propane C3H8 46 32 19 92 50 32 21 64 Butane C4H10...

Страница 22: ...or information on installing and removing the T DAT see Operating Instructions BA00115D 06 a0001221 en Fig 3 Data backup transmission with T DAT SAVE LOAD function Information on the LOAD and SAVE opt...

Страница 23: ...ou can enable programming by entering the private code factory setting 65 see DEFINE PRIVATE CODE function 23 The programming levels are disabled if you do not press the operating elements within 60 s...

Страница 24: ...nnot be modified 1 ACCESS CUSTOMER parameters can be modified MODBUS register Data type Access 2512 Integer read write ACCESS CODE COUNTER Description The number of times the private and service code...

Страница 25: ...ration Options 0 OFF 1 MASS FLOW 2 MASS FLOW IN 3 MASS FLOW BARGRAPH IN 4 TEMPERATURE 5 TOTALIZER 1 6 TOTALIZER 2 9 TAG NAME 10 OPERATING SYSTEM CONDITIONS 11 CORRECTED VOLUME FLOW 12 CORRECTED VOLUME...

Страница 26: ...r the decimal point as computed by the measuring device cannot always be displayed depending on this setting and the engineering unit In these instances an arrow appears on the display between the mea...

Страница 27: ...ility of the local display and its pixels Test sequence 1 Start the test by selecting ON 2 All pixels of the main line and additional line are darkened for minimum 0 75 seconds 3 The main line and add...

Страница 28: ...aggregated since measuring commenced appears on the display Total flow is represented by a floating point number consisting of max 7 digits Higher numerical values 9 999 999 can be viewed as overflows...

Страница 29: ...user 29 RESET TOTALIZER Description Resets the sum and overflow in the totalizer selected Options 0 NO 1 YES Factory setting NO MODBUS register TOTALIZER 1 TOTALIZER 2 Data type Access 2608 2808 Integ...

Страница 30: ...1 YES Factory setting NO MODBUS register Data type Access 2609 Integer read write FAILSAFE MODE Description For selecting the behavior of the totalizer in an alarm condition Options 0 STOP The totaliz...

Страница 31: ...ing OFF MODBUS register Data type Access 4301 Integer read write ACTIVE LEVEL Description Use this function to define whether the assigned switching function see ASSIGN STATUS INPUT function is trigge...

Страница 32: ...ata type Access 4306 Integer read write VALUE SIMULATION STATUS INPUT Prerequisite This function is only available if ON was selected in the SIMULATION STATUS INPUT function 32 Description Use this fu...

Страница 33: ...e this function for selecting the baud rate Options 0 1200 BAUD 1 2400 BAUD 2 4800 BAUD 3 9600 BAUD 4 19200 BAUD 5 38400 BAUD 6 57600 BAUD 7 115200 BAUD Factory setting 19200 BAUD MODBUS register Data...

Страница 34: ...vated and deactivated by means of a jumper on the I O module see Operating Instructions BA00115D 06 Display 0 OFF write access via MODBUS possible 1 ON write access via MODBUS blocked Factory setting...

Страница 35: ...nt number Factory setting 1 0130 bar a or 14 692 psi a country dependent 55 MODBUS register Data type Access 5270 Float read write PROCESS PRESSURE Description Use this function to display the pressur...

Страница 36: ...MASS 8 or UNIT CALORIFIC VALUE CORRECTED VOLUME 9 Display 5 digit floating point number MODBUS register Data type Access 5272 Float read GROSS CALORIFIC VALUE Prerequisite This function is only avail...

Страница 37: ...ing point number Factory setting 1 of full scale value MODBUS register Data type Access 5138 Float read write OFF VALUE LOW FLOW CUT OFF Description Use this function to enter the off value for low fl...

Страница 38: ...function RESET restores the factory calibration If the status input is appropriately configured the zero point adjustment can also be activated by using the status input see the ASSIGN STATUS INPUT fu...

Страница 39: ...orresponding Mole can be entered in any order within the mixture A gas constituent may have a value of 0 0 Mole within the mixture The option NOT USED is only a place holder for functions GAS TYPE 1 a...

Страница 40: ...t are used in the gas group User input 1 to 8 Factory setting 1 MODBUS register Data type Access 9717 Float read write GAS TYPE 1 Description Use this function to select gas type 1 Options 0 AIR 1 AMM...

Страница 41: ...the specified process conditions The correction factor is determined by the factory If the gas or process conditions change from the initial setting then the correction factor value will also need upd...

Страница 42: ...number of functions available here is dependent upon the setting in the function NUMBER OF GASES 40 Description Use this function to enter the Mole of the GAS TYPE 2 to 8 User input 000 00 to 100 00 F...

Страница 43: ...measurement 1 CANCEL The entered parameters are saved in the gas group but they are not used for flow measurement The gas group can be activated at a later time by returning to the group checking the...

Страница 44: ...a user specific calorific value Input display 5 digit floating point number Factory setting 0 0 The corresponding unit is taken from the UNIT CALORIFIC VALUE MASS 8 or UNIT CALORIFIC VALUE CORRECTED V...

Страница 45: ...escription Use this function to enter the reference combustion temperature of the gas This function is used to calculate the calorific value of the gas The unit is taken from the function UNIT TEMPERA...

Страница 46: ...put is appropriately configured the positive zero return function can also be activated by using the status input ASSIGN STATUS INPUT 31 Options 0 OFF 1 ON signal output is set to zero flow value temp...

Страница 47: ...sors t mass 65I Description Use this function to select the type of pipe Options 0 CIRCULAR 1 RECTANGULAR Factory setting CIRCULAR MODBUS register Data type Access 9603 Integer read write PIPE STANDAR...

Страница 48: ...YPE 47 and OTHER was selected in the function PIPE STANDARD 47 Description Use this function to enter a value for the outer diameter if the required pipe standard was not available in the function PIP...

Страница 49: ...er Data type Access 9606 Float read write INTERNAL WIDTH Prerequisite This function is only available for insertion sensors t mass 65I This function is only available if RECTANGULAR was selected in th...

Страница 50: ...nction UNIT LENGTH 10 For more information on insertion depth calculation please refer to the Operating Instructions BA00115D 06 Display 5 digit floating point number MODBUS register Data type Access...

Страница 51: ...could affect the safety requirements of the process If fault and notice messages are not be suppressed than a value of 0 seconds must be entered here User input 0 to 100 s in steps of one second Facto...

Страница 52: ...y be used for measuring to a certain extent while the simulation is in progress The setting is not saved if the power supply fails Options 0 OFF 1 MASS FLOW 2 CORRECTED VOLUME FLOW 3 TEMPERATURE 4 HEA...

Страница 53: ...NSDUCER SERIAL NUMBER Description The serial number of the transducer appears on the display MODBUS register Data type Access 7461 String 16 read SOFTWARE REVISION NUMBER S DAT Description Use this fu...

Страница 54: ...SOFTWARE REVISION NUMBER AMPLIFIER Description Use this function to view the software revision number of the amplifier board MODBUS register Data type Access 7039 String 16 read SOFTWARE REVISION NUMB...

Страница 55: ...Norway Norwegian Denmark English Poland Polish England English Portugal Portuguese Finland Finnish Sweden Swedish France French Switzerland German Germany German Singapore English Hungary English Spai...

Страница 56: ...1000 3200 00 320000 100 0 1500 7200 00 720000 100 0 Unit Unit Temperature C Length mm Density kg m Pressure bar a Reference density kg m Reference Pressure bar a Calorific Value Mass MJ kg Calorific V...

Страница 57: ...20 1220 2 00 2 20 02 2002 20 00 3 44 66 4466 20 00 4 82 50 8250 20 00 Nominal diameter Low flow cut off Full scale value Pulse value mm lb hr lb hr lb p 3 44 66 4466 20 00 4 82 50 8250 20 00 6 165 00...

Страница 58: ...e t mass 65 MODBUS RS485 58 Endress Hauser 21 3 3 Unit totalizer 1 and 2 28 21 3 4 Other Units Unit Unit Mass flow lb Corrected volume flow Sm Heat flow kBtu Unit Ref combustion temp F 45 Pipe standar...

Страница 59: ...50 FLOW DAMPING fct 46 FORMAT fct 26 G Gas pressure 19 Gas programming 17 GAS TYPE 1 fct 40 GAS TYPE 2 to 8 fct 42 GAS function group 39 GROSS CALORIFIC VALUE fct 36 H HANDLING TOTALIZER function grou...

Страница 60: ...ct 53 SIMULATION FAILSAFE MODE fct 52 SIMULATION MEASURAND fct 52 SIMULATION STATUS INPUT fct 32 SIMULATION SYSTEM function group 52 SOFTWARE REVISION NUMBER AMPLIFIER fct 54 SOFTWARE REVISION NUMBER...

Страница 61: ...__________________________________________ ______________________________________________________________________________________________________________ We hereby certify that this declaration is fil...

Страница 62: ...www endress com worldwide BA00116D 06 EN 13 10 71123867 FM SGML6 0 ProMoDo...