Commissioning
Proline Promass E 300 Modbus RS485
90
Hauser
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Failure mode
The
Pulse
option is selected in
the
Operating mode
parameter (→ 88) and a
process variable is selected in
the
Assign pulse output
Define output behavior in
alarm condition.
• Actual value
• No pulses
–
Invert output signal
–
Invert the output signal.
• No
• Yes
–
*
Visibility depends on order options or device settings
Configuring the frequency output
Navigation
"Setup" menu → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
Terminal number
Signal mode
Assign frequency output
Minimum frequency value
Maximum frequency value
Measuring value at minimum
frequency
Measuring value at maximum
frequency
Failure mode
Failure frequency
Invert output signal