Proline Promass E 300 Modbus RS485
Commissioning
Hauser
107
10.3.14 Configuring the partial filled pipe detection
The
Partial filled pipe detection
wizard guides you systematically through all parameters
that have to be set for configuring the monitoring of the pipe filling.
Navigation
"Setup" menu → Partially filled pipe detection
‣
Partially filled pipe detection
Assign process variable
Low value partial filled pipe detection
High value partial filled pipe detection
Response time part. filled pipe detect.
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User entry
Assign process variable
–
Select process variable for partially
filled pipe detection.
• Off
• Density
• Reference density
Low value partial filled pipe
detection
A process variable is selected in the
Assign process variable
parameter
Enter lower limit value for deactivating
partialy filled pipe detection.
Signed floating-point
number
High value partial filled pipe
detection
A process variable is selected in the
Assign process variable
parameter
Enter upper limit value for deactivating
partialy filled pipe detection.
Signed floating-point
number
Response time part. filled pipe
detect.
A process variable is selected in the
Assign process variable
parameter
Enter time before diagnostic message is
displayed for partially filled pipe
detection.
0 to 100 s