Proline Promass E 300 Modbus RS485
Commissioning
Hauser
123
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User entry /
User interface
Assign simulation process variable
–
Select a process variable for the
simulation process that is activated.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Target volume flow
• Carrier volume flow
• Target corrected volume
flow
• Carrier corrected volume
flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow alternative
• NSV flow
• NSV flow alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected volume
flow
• Water corrected volume
flow
• Temperature
• Concentration
Process variable value
A process variable is selected in the
Assign simulation process variable
Enter the simulation value for the
selected process variable.
Depends on the process
variable selected
Status input simulation 1 to n
–
Switch simulation of the status input on
and off.
• Off
• On
Input signal level 1 to n
In the
Status input simulation
parameter, the
On
option is selected.
Select the signal level for the simulation
of the status input.
• High
• Low
Current input 1 to n simulation
–
Switch simulation of the current input
on and off.
• Off
• On
Value current input 1 to n
In the
Current input 1 to n simulation
parameter, the
On
option is selected.
Enter the current value for simulation.
0 to 22.5 mA
Current output 1 to n simulation
–
Switch the simulation of the current
output on and off.
• Off
• On
Value current output 1 to n
In the
Current output 1 to n
simulation
parameter, the
On
option is
selected.
Enter the current value for simulation.
3.59 to 22.5 mA
Frequency output simulation 1 to n
In the
Operating mode
parameter, the
Frequency
option is selected.
Switch the simulation of the frequency
output on and off.
• Off
• On
Frequency value 1 to n
In the
Frequency output simulation
1 to n
parameter, the
On
option is
selected.
Enter the frequency value for the
simulation.
0.0 to 12
500.0 Hz