Proline Promass E 300 Modbus RS485
Commissioning
Hauser
121
Parameter overview with brief description
Parameter
Description
User entry
Define access code
Restrict write-access to parameters to protect the configuration
of the device against unintentional changes.
Max. 16-digit character string comprising
numbers, letters and special characters
Confirm access code
Confirm the entered access code.
Max. 16-digit character string comprising
numbers, letters and special characters
Using the parameter to reset the access code
Navigation
"Setup" menu → Advanced setup → Administration → Reset access code
‣
Reset access code
Operating time
Reset access code
Parameter overview with brief description
Parameter
Description
User interface / User entry
Operating time
Indicates how long the device has been in operation.
Days (d), hours (h), minutes (m) and seconds
(s)
Reset access code
Reset access code to factory settings.
For a reset code, contact your Hauser service
organization.
The reset code can only be entered via:
• Web browser
• DeviceCare, FieldCare (via service interface CDI-RJ45)
• Fieldbus
Character string comprising numbers, letters
and special characters
Using the parameter to reset the device
Navigation
"Setup" menu → Advanced setup → Administration
Parameter overview with brief description
Parameter
Description
Selection
Device reset
Reset the device configuration - either entirely or in part - to a
defined state.
• Cancel
• To delivery settings
• Restart device
• Restore S-DAT backup
*
Visibility depends on order options or device settings
10.5 Simulation
The
Simulation
submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).