Diagnostics and troubleshooting
Proline Promass E 300 Modbus RS485
154
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
882
Input signal
1. Check input configuration
2. Check external device or process
conditions
F
Alarm
910
Tubes not oscillating
1. Check electronic
2. Inspect sensor
F
Alarm
912
Medium inhomogeneous 1. Check process cond.
2. Increase system pressure
S
Warning
913
Medium unsuitable
1. Check process conditions
2. Check electronic modules or sensor
S
Warning
941
API temperature out of
specification
1. Check process temperature with
selected API commodity group
2. Check API related parameters
S
Warning
942
API density out of
specification
1. Check process density with selected
API commodity group
2. Check API related parameters
S
Warning
943
API pressure out of
specification
1. Check process pressure with
selected API commodity group
2. Check API related parameters
S
Warning
944
Monitoring failed
Check process conditions for Heartbeat
Monitoring
S
Warning
948
Oscillation damping too
high
Check process conditions
S
Warning
1)
Diagnostic behavior can be changed.
12.9 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 148
• Via "DeviceCare" operating tool → 148
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time