Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Control Totalizer 1 to n
A process variable is selected
in the
Assign process variable
Totalizer 1 to n
submenu.
Control totalizer value.
• Totalize
• Reset + hold
• hold
• Reset + totalize
• totalize
• Hold
–
Preset value 1 to n
A process variable is selected
in the
Assign process variable
Totalizer 1 to n
submenu.
Specify start value for totalizer.
Dependency
The unit of the selected
process variable is
specified for the totalizer
in the
Unit totalizer
parameter (→ 112).
Signed floating-point
number
Country-specific:
• 0 kg
• 0 lb
Reset all totalizers
–
Reset all totalizers to 0 and
start.
• Cancel
• Reset + totalize
–
*
Visibility depends on order options or device settings
11.6.1 Function scope of the "Control Totalizer" parameter
Options
Description
Totalize
The totalizer is started or continues running.
Reset + hold
The totaling process is stopped and the totalizer is reset to 0.
hold
The totaling process is stopped and the totalizer is set to its defined start value
from the
Preset value
parameter.
Reset + totalize
The totalizer is reset to 0 and the totaling process is restarted.
totalize
The totalizer is set to the defined start value from the
Preset value
parameter and
the totaling process is restarted.
Hold
Totalizing is stopped.
11.6.2 Function scope of the "Reset all totalizers" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
Reset + totalize
Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
11.7 Showing data logging
The
Extended HistoROM
application package must be enabled in the device (order
option) for the
Data logging
submenu to appear. This contains all the parameters for the
measured value history.
Data logging is also available via:
• Plant Asset Management Tool FieldCare → 66.
• Web browser