Commissioning
Proline Promass E 300 Modbus RS485
100
Hauser
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Switch-off delay
In the
Relay output function
parameter, the
Limit
option is
selected.
Define delay for the switch-off
of status output.
0.0 to 100.0 s
–
Switch-on value
The
Limit
option is selected in
the
Relay output function
parameter.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
• 0 kg/h
• 0 lb/min
Switch-on delay
In the
Relay output function
parameter, the
Limit
option is
selected.
Define delay for the switch-on
of status output.
0.0 to 100.0 s
–
Failure mode
–
Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
–
*
Visibility depends on order options or device settings
10.3.11 Configuring the double pulse output
The
Double pulse output
submenu guides the user systematically through all the
parameters that have to be set for configuring the double pulse output.
Navigation
"Setup" menu → Double pulse output
‣
Double pulse output
Master terminal number
Slave terminal number
Signal mode
Assign pulse output 1
Measuring mode
Value per pulse
Pulse width
Failure mode
Invert output signal