10 Technical data
PROline Promag 10
80
Hauser
10.1.7
Operating conditions
Installation
Installation instructions
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
Inlet and outlet runs
Inlet run: typically
≥
5 x DN
Outlet run: typically
≥
2 x DN
Length of connecting
cable
For the remote version the permissible cable length L
max
depends on the conductivity
of the medium
A minimum conductivity of 50
µ
S/cm is required for measuring demineralised water.
Environment
Ambient temperature
range
−
20...+60 °C (sensor, transmitter)
Please note the following points:
•
Install the device at a shady location. Avoid direct sunlight, particularly in warm
climatic regions.
•
The transmitter must be mounted separate from the sensor if both the ambient and
fluid temperatures are high (
→
“Fluid temperature”).
Storage temperature
–10...+50 °C (preferably at +20 °C)
Degree of protection
•
Standard: IP 67 (NEMA 4X) for transmitter and sensor
•
Optional: IP 68 (NEMA 6P) for remote version of Promag W and P sensor
Shock and vibration
resistance
Acceleration up to 2 g by analogy with IEC 68-2-6
CIP cleaning
Promag W: not possible
Promag P: possible (note max. temperature)
Promag H: possible (note max. temperature)
SIP cleaning
Promag W: not possible
Promag P: not possible
Promag H: possible (observe max. temperature)
Electromagnetic
compatibility (EMC)
To EN 61326
Emission: according to limit value for industrial environments documented in EN55011.