PROline Promag 10
10 Technical data
Hauser
87
10.1.9
Human interface
Display elements
•
Liquid crystal display: unilluminated, two-line, 16 characters per line
•
Display (operating mode) preconfigured: volume flow and totalizer status
•
1 totalizer
Operating elements
•
Local operation via three push buttons (–/+/ E)
Remote operation
Operation via HART protocol
10.1.10 Certificates and approvals
Ex approval
Information about currently available Ex versions (ATEX, FM, CSA, etc.) can be supplied
by your E+H Sales Centre on request. All explosion protection data are given in a
separate documentation which is available upon request.
Sanitary compatibility
Promag W:
No applicable approvals or certification
Promag P:
No applicable approvals or certification
Promag H:
•
3A approval and EHEDG-tested
•
Seals conform to FDA (apart from Kalrez seals)
Pressure measuring
device approval
Devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3
(3) of the EC Directive 97/23/EC (Pressure Equipment Directive). For larger nominal
diameters, certified flowmeters to Category III are optionally also available if necessary
(depends on fluid and operating pressure). All devices are applicable for all fluids and
instable gases on principle and have been designed and manufactured in accordance
to sound engineering practice.
CE approval
The measuring system is in conformity with the statutory requirements of the
EC Directives.
Hauser confirms successful testing of the device by affixing to it the CE mark.
Other standards and
guidelines
EN 60529:
Degrees of protection by housing (IP code)
EN 61010
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures.
EN 61326 (IEC 1326)
“Emmision to class A requirements”
Electromagnetic compatibility (EMC requirements)
ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related
Equipment - General Requirements.
Pollution degree 2, Installation Category II.
CAN/CSA-C22.2
No. 1010.1-92
Safety requirements for Electrical Equipment for Measurement and Control and Labora-
tory Use.
Pollution degree 2, Installation Category II