Mounting ground disks
Ground disks must also be mounted between the sensor and the pipe flange for
potential equalisation under certain application conditions, e.g. in the case of lined or
ungrounded piping (see Page 43 ff.). Ground disks can be ordered as separate
accessories from E+H (see Page 64).
"
Caution!
•
The length increases when ground disks (incl. seals) are used! Dimensions are found
in the Technical Information on Promag 10 (separate documents)
•
Additional seals must be mounted between the ground disk and the pipe flange.
1.
Place the ground disk and additional seal(s) between the instrument and the pipe
flange (see Fig. 17).
2.
Insert the bolts through the flange holes. Tighten the nuts so that they are still loose.
3.
Now rotate the ground disk as shown in Fig. 17 until the handle strikes the bolts.
This will centre the ground disk automatically.
4.
Now tighten the bolts to the required torque (see Page 24 ff.).
5.
Connect the ground disk to ground
Fig. 19:
Assembly with ground disks (Promag P, DN 25…300)