PROline Promag 10
6 Commissioning
Hauser
61
6.2.2
Brief “commissioning” guide
These brief instructions guide you systematically through the setup procedure for all the
major device functions that have to be configured for standard measuring operation.
Configure display
Display language
Display contrast
Number of decimal
places
▼
Select engineering units
Volume flow
Totalizer
▼
Configure outputs
Current output
Pulse/status output
Current span
Operating mode
Full scale value
Pulse value
Pulse width
or
Assign status output
Switch-on point
▼
More complex applications
require other functions be programmed. Notes on this can be found on the following
pages:
Functions at a glance
Index
Operating matrix
▼
For optimum measurement results
Low flow cut off
Empty pipe detection