10 Technical data
PROline Promag 10
82
Hauser
Medium pressure range
(nominal pressure)
Promag W
Promag P
Promag H
The permissible nominal pressure depends on the process connection and seal:
•
40 bar: flange, weld nipple (with O-ring seal)
•
16 bar: all other process connections
Pressure tightness
(measuring tube lining)
DIN 2501
PN 10 (DN 200...600), PN 16 (DN 65...150), PN 40 (DN 25...50)
ANSI B16.5
Class 150 (1...24")
JIS B2238
10K (DN 65...600), 20K (DN 25...50)
DIN 2501
PN 10 (DN 200...300), PN 16 (DN 65...150), PN 40 (DN 25...150)
ANSI B16.5
Class 150 (1...12")
JIS B2238
10K (DN 65...300), 20K (DN 25...50)
Promag W
nominal diameter
Measuring
tube lining
Pressure tightness, measuring tube lining
Limit values for absolute pressure [mbar] with different fluid
temperatures
[mm]
[inch]
25 °C
60°C
80 °C
100 °C
130 °C
150 °C
180 °C
25...600
1...24"
Polyurethane
0
0
−
−
−
−
−
65...600
1...24"
Hard rubber
0
0
0
−
−
−
−
Promag P
nominal
diameter
Measuring tube
lining
Pressure tightness, measuring tube lining
Limit values for absolute pressure [mbar] with different fluid
temperatures
[mm]
[inch]
25 °C
80 °C
100 °C
130 °C
25
1"
PTFE
0
0
0
100
32
−
PTFE
0
0
0
100
40
1 1/2"
PTFE
0
0
0
100
50
2"
PTFE
0
0
0
100
65
−
PTFE
0
*
40
130
80
3"
PTFE
0
*
40
130
100
4"
PTFE
0
*
135
170
125
−
PTFE
135
*
240
385
150
6"
PTFE
135
*
240
385
200
8"
PTFE
200
*
290
410
250
10"
PTFE
330
*
400
530
300
12"
PTFE
400
*
500
630
* No value can be specified.
Promag H
nominal diameter
Measuring tube
lining
Pressure tightness, measuring tube lining
Limit values for absolute pressure [mbar] with different
fluid temperatures
[mm]
[inch]
25 °C
80 °C
100 °C
130 °C
150 °C
180 °C
2 ...100
1/12...4"
PFA
0
0
0
0
0
0