6 Commissioning
PROline Promag 10
62
Hauser
6.2.3
Commissioning after installing a new electronics board
If a new electronics board has been installed, the sensor-specific data must be entered
manually. After start-up, the device checks whether a serial number is available. If this
is not the case, the following setup is started.
Setup "Commissioning"
!
Note!
•
The setup can no longer be called up once a serial number has been entered and
stored. If a parameter is entered incorrectly during the setup, this must be corrected
in the relevant function via the function matrix.
•
The appropriate information is documented on the nameplate of the sensor and in the
interior of the housing cover (see Fig. 2, Page 8).
Fig. 38:
Setup "Commissioning". Starts after installation of a new electronics board if no serial number is
present.
-
+
HAUSER
E
ESC
Language
Nominal
Diameter
K-Factor
Zero
Point
EPD
Electrode
HOME-POSITION
Serial
Number
Equal?
No
Yes
Setup
complete
HOME
Verif. S#
F06-
10
xx
xx
xx
-19-
x
x
-x
x
-en-
000