Commissioning
Micropilot S FMR533 4 to 20 mA HART
50
Hauser
Function "history reset" (009)
By this function a history reset of the device is performed, i.e. the correspondance table between
level an index values is deleted. A new correspondance table will be filled and stored after the
history reset, cf.
"
Caution!
A history reset must be performed after:
• first installation or
• change of basic setup or
• change of the installation situation.
In this case also effect a reset of the dip table in function "
dip table mode
"
(033)
.
⇒
E HAUSER
E
+
–
⇒
⇓
After 3 s, the following message appears
E HAUSER