Micropilot S FMR533 4 to 20 mA HART
Commissioning
Hauser
65
6.6.1
Signal analysis via envelope curve
After the basic setup, an evaluation of the measurement using the envelope curve is recommended.
MicropilotM-en-306
!
Note!
If the level of echo is very weak or there is a heavy interference echo, an orientation of the
Micropilot can help optimise the measurement (increase of the useful echo/reduction of the
interference echo).
6.6.2
User-specific applications (operation)
For details of setting the parameters of user-specific applications, see separate documentation
BA00217F/00/EN "Description of Instrument Functions" on the enclosed CD-ROM.
6.7
Mounting calibration with the
Hauser operating program
To carry out the mounting calibration with the operating program, proceed as follows:
• Start the operating program and establish a connection.
• Select the "
mounting calibr
." function group in the navigation bar.
The following display appears on the screen:
Mounting calibration step 1/2:
• auto correction
• pipe diam. corr.