Micropilot M FMR 240 with PROFIBUS PA
Mounting
E Hauser
25
3.4.5
Installation FMR 240 with Wave Guide antenna
Optimum mounting position
L00-FMR240xx-17-00-00-de-003
Standard installation
For installations in a stilling well, follow the engineering hints on Page 14 and note the
following points:
•
Can be mounted in a tank or bypass.
•
No alignment is required.
•
After mounting, the housing can be turned 350° in order to simplify access to the
display and the terminal compartment.
•
Only suitable for media with low viscosity (
≤
500 cst), without the tendency for
build-up in the Wave Guide antenna.
•
No minimum distance is required between the Wave Guide antenna and the tank
bottom for measuring-related purposes.
!
Note!
In the event of horizontal stress (e.g. agitators) , mechanical support is required or
provide the Wave Guide antenna with a protective pipe.