Micropilot M FMR 240 with PROFIBUS PA
Commissioning
E Hauser
65
Function "full calibr." (006)
This function is used to enter the distance from the minimum level to the maximum level
(=span).
L00-FMR2xxxx-14-00-06-en-009
!
Note!
If
bypass
or
stilling well
was selected in the "
tank shape
"
(002)
function, the pipe
diameter is requested in the following step.
In principle, it is possible to measure up to the tip of the antenna. However, due to
considerations regarding corrosion and build-up, the end of the measuring range
should not be chosen any closer than 50 mm (2”) to the tip of the antenna.
⇒