Operation
Micropilot M FMR 240 with PROFIBUS PA
46
E Hauser
Status codes
The status codes comprise one byte and have got the following meaning:
If a status other than "GOOD" is sent to the device, the display indicates an error.
Status-
Code
Device status
Significance
Primary value
Secondary
value
0C Hex
BAD
device error
X
0F Hex
BAD
device error
X
1F Hex
BAD
out-of-service (target mode)
X
40 Hex
UNCERTAIN
non-specific (simulation)
X
47 Hex
UNCERTAIN
last usable value
(Fail-safe-Mode aktiv)
X
4B Hex
UNCERTAIN
Substitute set
(fail-Safe mode active)
X
4F Hex
UNCERTAIN
initial value
(fail-Safe mode active)
X
5C Hex
UNCERTAIN
Configuration error
(limits not set correctly)
X
80 Hex
GOOD
OK
X
X
84 Hex
GOOD
Active block alarm
(static revision counter incremented)
X
89 Hex
GOOD
LOW_LIM (alarm active)
X
8A Hex
GOOD
HI_LIM (alarm active)
X
8D Hex
GOOD
LOW_LOW_LIM (alarm active)
X
8E Hex
GOOD
HI_HI_LIM (alarm active)
X